ERROR 509
The NX Tester has Autolearned a wire harness, but there is not sufficient storage space on the NX Tester
to save the program file.
Corrective Action
Delete some files on the NX Tester to make room, then execute the Autolearn again.
ERROR 510
The Autolearn template contains switches that conflict with connections found in the Autolearn scan. It
is illegal to have continuity between one side of a switch and a test point other than the other terminal of
the switch. Nor is it legal to have a resistor across a switch.
Corrective Action
Since the Autolearn template is incompatible with the wire harness configuration, the template
should be modified accordingly – or, the wire harness is possible incorrectly wired.
ERROR 601
The NX Tester has failed diagnostics of the serial ports.
Corrective Action
Verify that the Dynalab transfer cable was used to connect the two serial ports being tested. If the
proper cable was used, contact Dynalab.
ERROR 602
The internal calibration process has failed.
Corrective Action
: Contact Dynalab
ERROR 603
The firmware and CPU board are of incompatible versions.
Corrective Action
: Contact Dynalab
ERROR 606
The internal 3.3V supply voltage on the multifunction board is higher than the acceptable upper limit.
Corrective Action
: Contact Dynalab
ERROR 607
The internal 3.3V supply voltage on the multifunction board is lower than the acceptable lower limit.
Corrective Action
: Contact Dynalab
ERROR 608
The internal 5V analog supply voltage on the CPU board is higher than the acceptable upper limit.
Corrective Action
: Contact Dynalab
ERROR 609
The internal 5V analog supply voltage on the CPU board is lower than the acceptable lower limit.