Dynalab Test Systems
NX Tester User’s Guide
35
Copyright 2011
Revised 04/27/2011
ERROR 636
The current level supplied by the internal 5uA current source on the CPU board is higher than the
acceptable upper limit.
Corrective Action
: Contact Dynalab
ERROR 637
The current level supplied by the internal 5uA current source on the CPU board is below the acceptable
lower limit.
Corrective Action
: Contact Dynalab
ERROR 638
The window voltage for the capacitor compare circuit is higher than the acceptable upper limit.
Corrective Action
: Contact Dynalab
ERROR 639
The window voltage for the capacitor compare circuit is below the acceptable lower limit.
Corrective Action
: Contact Dynalab
ERROR 640
The gain of the instrument amplifier is higher than the acceptable upper limit.
Corrective Action
: Contact Dynalab
ERROR 641
The gain of the instrument amplifier is below the acceptable lower limit.
Corrective Action
: Contact Dynalab
ERROR 702
A program references test points that are not present on the tester. For instance, a program that
references 256 test points will not operate on a NX Tester equipped with only 128 test points.
Corrective Action
Ensure that the NX Tester is equipped with at least as many test points as are referenced by the NX
Editor program. In the case of an NX Pro+ Tester, make sure that the dip switches are properly set
as illustrated in a previous section of this document.
ERROR 704
The file fails the Audit - and was found to have errors that would prevent the file from executing
properly.
Corrective Action
Open the file with the NX Editor, execute the Audit and correct the errors.
ERROR 705
The test program has a checksum specified that does not match the calculated checksum value.