ERROR 805
The Workflow contains a TEST cell that references phase with no defined connections.
Corrective Action
Change the test program so that all phases referenced by TEST Workflow items have at least one
defined connection.
ERROR 806
A Decide Workflow Item contains either a string operation on an integer variable or an integer operation
on a string variable.
Corrective Action
Use the NX Editor to modify the program’s Workflow. Change the Decide Workflow item so that
comparisons are made on like variables. Both variables in the comparison must be strings or both
variables must be counters.
ERROR 807
The Workflow referenced a Multiport port that does not exist. This could happen if there is no serial
port configured as a multi-port, the specified multi-port is not present, or the specified multi-port is
outside the range that is supported by the multi-port module.
Corrective Action
Make sure that a Multiport unit with the correct number of ports is connected to the NX Tester, and
that the associated serial port is properly configured for Multiport use.
ALL OTHER ERRORS
For any error not specifically listed above, please contact Dynalab.
Service and Support
For problems or questions requiring service or support, please contact Dynalab in one of the following
ways:
•
Send an email, describing the problem to:
•
Call 614-729-6550
•
FAX a description of the problem to 614-729-6551
Dynalab’s technical support engineers will work with you to resolve any problems you have.