Error
Possible cause
Remedy
Defective membrane drying unit
❯
Replace the membrane drying
unit.
❯
Inform a Service Technician.
Cup seal worn at the piston or
defective
❯
Replace the cup seal or the
entire piston.
Water dripping from air con-
sumers
Maintenance work not carried
out regularly (without membrane
drying unit)
❯
Regularly drain the condensa-
tion water from the pressure
tank, see "9.4 Draining the
condensation water"
Defective membrane drying unit
❯
Inform a service technician.
Working cycles of the com-
pressor are very short, even if
only small amounts of air are
removed
Condensed water in the tank
❯
Drain off condensed water
❯
The dry air units of compres-
sors fitted with dry air units
need to be checked and
replaced, if applicable.
Troubleshooting
42
9000-610-60/02 1909V007
EN