4.5
Tornado 2
Electrical data
5280-01
5280-03
5285-01
Nominal voltage
V
230
230
Mains frequency
Hz
50
60
50
60
Nominal current at 8 bar (0.8 MPa)
A
7.6
6.9
7.6
6.9
Motor protection
Motor winding over-
heat protector
Motor winding over-
heat protector
Speed
rpm
1392
1662
1392
1662
Type of protection
IP 24
IP 24
Mains fuses *
A
16
16
Max. permissible mains impedance in
accordance with EN 61000
‑
3
‑
11
W
(0.276 + j0.172)
(0.276 + j0.172)
*
Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1
General technical data
Pressure tank volume
l
20
20
Suction power, approx.
l/min
260
315
260
315
Delivery at 5 bar (0.5 MPa)
l/min
124
140
124
140
Pressure build-up phase 0 - 7.5 bar (0 -
0.75 MPa) c.
s
53
-
53
-
Duty cycle
%
100
100
Start-up pressure
bar (MPa)
6 (0.6)
6 (0.6)
Cut-off pressure
bar (MPa)
7.8 (0.78)
7.8 (0.78)
Cut-off pressure, max. adjustable
bar (MPa)
9.5 (0.95)
9.5 (0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
10 (1)
Dimensions (H x W x D) *
cm
65 x 48 x 41
84 x 63 x 53
Weight
kg
38
51
Noise level **
dB(A)
68
-
56
61
*
Values without accessories and add-on parts
**
ISO 3746.
Filter mesh size
Compressor unit intake filter
µm
3
Ambient conditions during storage and transport
Temperature
°C
-10 to +55
Relative humidity
%
max. 95
Ambient conditions during operation
Temperature
°C
+10 to +40
Product description
16
9000-610-60/02 1909V007
EN
Summary of Contents for Tornado 1
Page 2: ......
Page 13: ...Classification Medical devices class IIa Product description 9000 610 60 02 1909V007 11 EN ...
Page 46: ......