Assembly
6
Requirements
The unit must not be set up or operated
within the vicinity of the patients (within a
radius of 1.5 m).
The unit can be installed either at the same level
as the surgery room or on a floor below (e.g. cel-
lar).
Due of the amount of noise generated, we rec-
ommend that the unit is installed in an adjoining
room.
The pipes provided on-site must at least meet
the country-specific requirements for drinking
water.
Further information can be found in our
separate planning information leaflet for
compressed air.
6.1
Installation/setup room
The room chosen for set up must fulfil the follow-
ing requirements:
– Closed, dry, well-ventilated room
– Should not be a room made for another pur-
pose (e. g. boiler room or wet cell)
– If the unit is installed in a machine room, e.g. in
an adjoining room or cellar, the requirements
set out in ISO
‑
TS 22595 must be complied
with.
NOTICE
Risk of overheating due to insufficient
ventilation
The units generates heat. Possibility of
heat damage and/or reduced service life
of the unit.
❯
Do not cover the unit.
❯
Install a fan for auxiliary ventilation in
rooms where ambient temperatures
exceed
³
40 °C while the unit is in
operation.
≥ 40 °C
6.2
Setup
The following conditions must be taken into
account for installation:
The air is filtered when it is sucked in. This
does not alter the composition of the air.
For this reason it is important to keep the
sucked-in air free of harmful substances
(e.g. do not suck in exhaust gases or
contaminated exhaust air).
– Clean, level and sufficiently stable subsurface
(note the weight of the unit).
– Type plate easy to read.
– Unit easy to access for operation and mainte-
nance.
– Easy-to-access power outlet to which the unit
is connected.
– Maintain sufficient distance to the wall (at least
20 cm).
– The compressed air pipe should be routed as
closely as possible to the place of installation
(note the length of the hose supplied).
6.3
Information about electrical
connections
❯
Ensure that electrical connections to the mains
power supply are carried out in accordance
with current valid national and local regulations
and standards governing the installation of low
voltage units in medical facilities.
❯
Install an all-pole disconnect switch with a con-
tact opening width of at least 3 mm in the elec-
trical connection to the mains power supply.
❯
Observe the current consumption of the
devices that are to be connected.
Assembly
9000-610-60/02 1909V007
23
EN