Secure the mains connector using the cable tie
provided.
❯
Connect the mains plug to an earthed power
outlet.
9
Commissioning
❯
Before taking the unit into operation, check for
any damage. Damaged units must not be
used.
In many countries technical medical prod-
ucts and electrical devices are subject to
regular checks at set intervals. The owner
must be instructed accordingly.
❯
Turn on the unit power switch or the main
surgery switch.
❯
Carry out an electrical safety check in accor-
dance with applicable local regulations (e.g. the
German Ordinance on the Installation, Opera-
tion and Use of Medical Devices / Medizinpro-
dukte-Betreiberverordnung) and record the
results as appropriate, e.g. in the technical log
book.
9.1
Check the motor protection
switch
In the case of compressors that are operated at
100 – 127 V, the motor protection switch must
be checked and adjusted in the event of devia-
tions. The motor protection switch is combined
with the pressure switch in a shared unit. It was
set to the recommended setting in the factory
(see "4 Technical data").
Compressors that are operated at 230 V have a
motor winding protector instead of a motor pro-
tection switch, which cannot be adjusted.
❯
Switch on the unit at the pressure switch by
turning it to the position "I AUTO".
Assembly
26
9000-610-60/02 1909V007
EN