Pressure tank
❯
At maximum tank pressure, slowly open the
condensate drain valve.
❯
Close the condensate drain valve as soon as
all of the accumulated condensation water has
been blown out.
10 Adjustment options
10.1 Adjusting the pressure switch
WARNING
Risk of explosion of the pressure ves-
sel
The pressure vessels used in the com-
pressors are designed to withstand
continuous pressure changes of 2 bar
and can be used continuously under
these pressure changes.
❯
For load changes > 2 bar (max. per-
missible: 3 bar), comply with the maxi-
mum load change cycles specified in
the operating instructions of the pres-
sure vessel.
DANGER
Exposed live parts
Risk of electric shock due to live parts
❯
Disconnect all power from the device.
❯
Use insulated tools.
❯
Do not touch live parts.
The cut-off pressure must be at least 0.5
bar (0.05 hPa) below the maximum pres-
sure of 10 bar (1 hPa) of the safety valve.
Otherwise the safety valve can open too
early, which will prevent the compressor
unit from attaining the cut-off pressure, as
a result of which it will run continuously.
The maximum permitted pressure is
marked by a red line on the attached
pressure gauge.
Adjustment at 230 V
If the read-off values differ from the factory set-
tings or if other settings are required, the cut-off
pressure of the compressor can be adjusted at
the adjusting screw on the pressure switch. The
pressure difference Δp cannot be altered.
❯
Take off the pressure switch cover.
Assembly
28
9000-610-60/02 1909V007
EN