
4.4
Tornado 1 with membrane drying unit
Electrical data
5182-02
Nominal voltage
V
100 - 110
100 - 127
Mains frequency
Hz
50
60
Nominal current at 8 bar (0.8 MPa)
A
8.6 - 9.0
9.1 - 8.0
Motor protection switch, recommended
settings
A
8.6 - 10
9.1 - 9.1
Speed
rpm
1400 - 1420
1655 - 1720
Type of protection
IP 24
Mains fuses *
A
16
*
Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1
General technical data
Pressure tank volume
l
20
Suction power, approx.
l/min
130
155
Delivery at 5 bar (0.5 MPa)
l/min
60
70
Pressure build-up phase 0 - 7.5 bar (0 -
0.75 MPa) c.
s
133
-
Duty cycle
%
100
Start-up pressure
bar (MPa)
6 (0.6)
Cut-off pressure
bar (MPa)
7.8 (0.78)
Cut-off pressure, max. adjustable
bar (MPa)
9.5 (0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
Pressure dew point at 7 bar (0.7 MPa) *
°C
£
+5
Dimensions (H x W x D) **
cm
65 x 51 x 47
Weight
kg
36
Noise level ***
dB(A)
64
-
*
Value determined at an ambient temperature of +40 °C
**
Values without accessories and add-on parts
*** In accordance with ISO 3746
Filter mesh size
Compressor unit intake filter
µm
3
Fine filter for membrane drying unit
µm
3
Sterile filter for membrane drying unit
µm
0.01
Sintered filter for membrane drying unit
µm
35
Ambient conditions during storage and transport
Temperature
°C
-10 to +55
Product description
14
9000-610-60/02 1909V007
EN
Summary of Contents for Tornado 1
Page 2: ......
Page 13: ...Classification Medical devices class IIa Product description 9000 610 60 02 1909V007 11 EN ...
Page 46: ......
Page 47: ......