5
Operation
5.1
Unit without membrane-dry-
ing unit
1
2
3
4
5
6
7
8
9
11
10
1
Compressor unit
2
Air intake filter
3
Carry handles
4
On/off switch
5
Pressure switch
6
Pressure gauge/display
7
Pressure tank
8
Compressed air connection (quick release
coupling)
9
Mains connection
10 Condensate drain valve
11 Safety valve
NOTICE
Risk of corrosion of the unit
Moisture can lead to premature corro-
sion.
❯
Retrofit a membrane drying unit.
The compressor unit draws in atmospheric air
and compresses it without oil. The oil-free com-
pressed air is then transported directly to the
pressure tank. The oil-free and hygienic air is
made available to the consumers in the pressure
tank.
5.2
Unit with membrane-drying
unit
1
2
3
4
5
6
7
8
9
13
12
11
10
14
15
1
Compressor unit
2
Air intake filter
3
Carry handles
4
On/off switch
5
Pressure switch
6
Pressure gauge/display
7
Pressure tank
8
Compressed air connection (quick release
coupling)
9
Mains connection
10 Condensate drain valve
11 Safety valve
12 Collector tray
13 Automatic/manual condensate drain valve,
membrane drying unit
14 Sinter filter for membrane drying unit
15 Fine filter or sterile filter, membrane drying
unit
The compressor unit draws in atmospheric air
and compresses it without oil. It then transports
the oil-free compressed air to the membrane dry-
ing unit. The cooler and the membrane dryer
extract moisture from the compressed air. The
oil-free, hygienic and dry air is stored in the pres-
sure tank ready for use in connected devices.
Product description
22
9000-610-60/02 1909V007
EN
Summary of Contents for Tornado 1
Page 2: ......
Page 13: ...Classification Medical devices class IIa Product description 9000 610 60 02 1909V007 11 EN ...