4
Technical data
4.1
Tornado 1
Electrical data
5180-01
5180-03
5185-01
Nominal voltage
V
230
230
Mains frequency
Hz
50
60
50
60
Nominal current at 8 bar (0.8 MPa)
A
3.7
4.4
3.7
4.4
Motor protection
Motor winding over-
heat protector
Motor winding over-
heat protector
Speed
rpm
1340
1560
1340
1560
Type of protection
IP 24
IP 24
Mains fuses *
A
10
10
*
Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1
General technical data
Pressure tank volume
l
20
20
Suction power, approx.
l/min
130
155
130
155
Delivery at 5 bar (0.5 MPa)
l/min
67
77
67
77
Pressure build-up phase 0 - 7.5 bar (0 -
0.75 MPa) c.
s
100
-
100
-
Duty cycle
%
100
100
Start-up pressure
bar (MPa)
6 (0.6)
6 (0.6)
Cut-off pressure
bar (MPa)
7.8 (0.78)
7.8 (0.78)
Cut-off pressure, max. adjustable
bar (MPa)
9.5 ( 0.95)
9.5 ( 0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
10 (1)
Dimensions (H x W x D) *
cm
66 x 48 x 41
84 x 63 x 53
Weight
kg
31
44
Noise level **
dB(A)
64
-
55
57
*
Values without accessories and add-on parts
**
In accordance with ISO 3746.
Filter mesh size
Compressor unit intake filter
µm
3
Ambient conditions during storage and transport
Temperature
°C
-10 to +55
Relative humidity
%
max. 95