7
Transport
WARNING
Risk of explosion of the pressure tank
and pressure hoses
❯
The pressure tank and the pressure
hoses must be vented before they are
stored or transported.
❯
Protect the unit against moisture, dirt and
extreme temperatures during transport (refer to
the section on "Ambient conditions").
❯
Always make sure that the condensate collec-
tor chamber is empty before transporting the
unit ("14.1 Taking the unit out of use").
❯
Always transport the unit in an upright position.
❯
Only transport the unit using the transport han-
dles provided.
8
Installation
8.1
Remove the transport locks
For safe transport, the unit is secured with foam
padding blocks and a transport strap.
❯
Cut and remove the transport strap.
❯
Remove the foam padding blocks.
2
1
1
Transport strap
2
Foam padding blocks
8.2
Establishing the compressed
air connection
The supplied flexible pressure hose
between the pipe system and the com-
pressor prevents vibrations from being
transmitted and thus reduces noise. This
ensures safe and reliable operation.
❯
Connect the pre-assembled connecting sleeve
of the pressure hose to the quick release cou-
pling.
❯
Measure the required length of the pressure
hose and shorten if necessary.
❯
Press a fitting hose connector (not included in
the scope of delivery) onto the pressure hose
(internal diameter 10 mm) and secure it with a
hose clip.
❯
Connect the connecting sleeve of the pressure
hose to the compressed air tube.
Assembly
24
9000-610-60/02 1909V007
EN