8.3
Place a collector tray under-
neath
During operation, condensation water on the unit
is continuously separated and automatically
drained. In order to prevent water damage due to
drained condensation, it is collected in the collec-
tor tray.
As an option, the condensation can also
be drained through a hose into the waste
water system. Always comply with appli-
cable national regulations for waste water
systems.
❯
Place a collector tray under the condensate
separator or the membrane drying unit
(depending on type).
8.4
Electrical connections
Safety when making electrical connections
The unit has no main power switch. For
this reason it is important that the unit is
be set up in such a way that the plug can
be easily accessed and unplugged if
required.
❯
The unit must only be connected to a correctly
installed power outlet.
❯
Make sure that none of the electrical cables
leading to the unit are under any mechanical
tension.
❯
Before initial start-up check that the mains sup-
ply voltage and the voltage stated on the type
plate match (see also "4. Technical data").
Establishing the electrical connections
DANGER
Risk of electric shock due to defective
mains cable
❯
Mains cables must not be allowed to
come into contact with any hot sur-
faces on the unit.
❯
Connect the connector of the mains cable to
the corresponding mains plug at the pressure
switch.
Assembly
9000-610-60/02 1909V007
25
EN