4.6
Tornado 2 with membrane drying unit
Electrical data
5282-01
5282-03
5282100029
5286-01
Rated voltage
V
230
230
Mains frequency
Hz
50
60
50
60
Nominal current at 8 bar (0.8 MPa)
A
7.7
7.0
7.7
7.0
Motor protection
Motor winding over-
heat protector
Motor winding over-
heat protector
Speed
rpm
1392
1662
1392
1662
Type of protection
IP 24
IP 24
Mains fuses *
A
16
16
Max. permissible mains impedance in
accordance with EN 61000
‑
3
‑
11
W
(0.276 + j0.172)
(0.276 + j0.172)
*
Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1
General technical data
Pressure tank volume
l
20
20
Suction power, approx.
l/min
260
315
260
315
Delivery at 5 bar (0.5 MPa)
l/min
110
126
110
126
Pressure build-up phase 0 - 7.5 bar (0 -
0.75 MPa) c.
s
73
-
73
-
Duty cycle
%
100
100
Start-up pressure
bar (MPa)
6 (0.6)
6 (0.6)
Cut-off pressure
bar (MPa)
7.8 (0.78)
7.8 (0.78)
Cut-off pressure, max. adjustable
bar (MPa)
9.5 (0.95)
9.5 (0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
10 (1)
Pressure dew point at 7 bar (0.7 MPa) *
°C
£
+5
£
+5
Dimensions (H x W x D) **
cm
65 x 49 x 47
84 x 63 x 60
Weight
kg
44
57
Noise level ***
dB(A)
68
-
59
62
*
Value determined at an ambient temperature of +40 °C
**
Values without accessories and add-on parts
*** In accordance with ISO 3746.
Filter mesh size
Compressor unit intake filter
µm
3
Fine filter for membrane drying unit
µm
3
Sterile filter for membrane drying unit
µm
0.01
Sintered filter for membrane drying unit
µm
35
Product description
18
9000-610-60/02 1909V007
EN