13.2 Replacing the air intake filter
Units without a noise reduction hood
❯
Switch off the compressor at the pressure
switch.
❯
Disconnect all power from the device.
❯
Release the filter cover by rotating it
anti-
clockwise
and then take it off.
❯
Remove the air intake filter.
❯
Insert a new air intake filter.
❯
Place the filter cover in position and lock it by
turning it
clockwise
.
Units with a noise reduction hood
❯
Switch off the compressor at the rotary switch
of the pressure switch.
❯
Disconnect all power from the device.
❯
Take off the noise reducing hood and the foam
filter cover.
2
1
❯
Release the filter cover by rotating it
anti-
clockwise
and then take it off.
2
1
❯
Remove the air intake filter.
❯
Insert a new air intake filter.
❯
Place the filter cover in position and lock it by
turning it
clockwise
.
❯
Fit the the foam filter cover and the noise
reducing hood.
13.3 Replacing the filter of the
membrane drying unit
Fine/sterile filter
❯
Switch off the unit.
❯
Disconnect all power from the device.
❯
Unscrew and remove the filter cover.
❯
Remove the fine/sterile filter.
❯
Insert the new fine/sterile filter.
Usage
38
9000-610-60/02 1909V007
EN