Verify the state of the linear actuator: if during the checks listed in step C the linear actuator has not
moved the calibration door, check the actuator’s connection to the Read NIR’s electronic board (open
the door on the side of the head unscrewing the 4 M4 self-locking nuts and check the 4-wired-
connector).
If during the checks listed in step C the calibration door does not move from one side to the other of
the reading window, check that there are not mechanical obstacles.
f)
Verify the optical fiber: execute the analysis of the component. If AGRINIR™ continues showing the
message “ANALYSIS ERROR!”, check that the extremity of the optical fiber inside the Read NIR
reading head’s window is clean.
WARNING!
The optical fiber is directly connected inside the NIR optical sensor and shows a little
glass fiber inside, that’s why:
•
THE NIR GROUP SHOULD NOT BE OPENED
(except for the lamp cover as indicated in
step D and E);
•
in case of faults, get in touch with
Assistance Service.
8.2
Replacement of the Printer’s Paper
a)
Turn the AGRINIR™ off.
b)
Open the front panel of the printer by pressing the central green key (it lights up when the AgriNIR™
system is switched on).
c)
Pull out the empty paper plastic support.
d)
Insert the new
THERMIC PAPER
roll, keeping attention to
insert correctly the paper through the printer site.
e)
Hook again the front panel of the printer, exerting a light pressure on the points show in the picture
below :
f)
Turn the AGRINIR™ on.
g)
Let the paper advance pushing <FEED>
on the printer. If the paper is blocked, check that it has
been inserted correctly.