TROUBLESHOOTING
1.
Verify if the CHECK-CELL is properly inserted inside the fodder box. Please see the
pictures at page 23 point 1.
2.
Verify if the fodder box with CHECK-CELL is properly inserted inside the AgriNIR™
analyzer box. Please see the pictures at page 24 point 2.
3.
Verify the cleaning of the CHECK-CELL glass or the ReadNIR glass inside the NIR
group on the bottom side.
4.
Verify if the CHECK-CELL glass or the ReadNIR glass is broken, otherwise contact
the
Customer Service for the spare parts.
5.
Verify if the lamp inside the NIR group is not on its correct position or its time life is
expiring (1 year); in second case, please contact the
Customer
Service for the spare part.
6.
Verify if the optic fiber terminal (visible inside the NIR group through the glass window
on the bottom side) is properly inserted on its brass fitting, otherwise if it's moved
please contact the
Customer Service.
7.
If it's none of the previous cases, please save the optical spectra on USB memory
key and contact the
Customer Service.