Use AgriNIR™ Analyzer Rev. A11
E
NG
L
IS
H
53
6.17 Deleting the initial message of FULL MEMORY
After a certain number of analysis (approx. 60-70) performed with the AgriNIR™ system, the following
message may appear when switching the system on :
When switching the system on, this message is
displayed to warn that the space of the internal
memory for the Analysis Execution Log is
reaching its full capacity.
NOTE.
This section of internal memory is NOT accessible to the user and may NOT be deleted with the
deleting function
, as this function refers only to the memory area for the analysis data performed.
To delete the section of internal memory for the Analysis Execution Log follow this procedure :
When switching the system on, when the
message “System Initialization” is displayed,
press <ENTER> for a few seconds until the
SYSTEM CONFIGU-RATION environment
appears.
Access the configuration menu 6 :
“Setup protected by password”