Use AgriNIR™ Analyzer Rev. A11
38
At the end of the analysis executions (single or multiple) the AgriNIR™ system shows the results :
On display it is shown a table that reports the prediction values of NIR chemical parameters obtained
from the analysis performed on the sample in the fodder box.
The values of the chemical parameters are always expressed in % and in 2 columns :
•
in the left column are the
prediction values expressed on AsIs
, that is to say the total %
present in the sample just analyzed;
•
in the right column are the
prediction values expressed on the DRY MATTER
, that tot o say on
the sample part without water.
NOTE.
In case of
MULTIPLE ANALYSIS
the AgriNIR™ system does not show the results after every
single analysis performed but only at the end of the execution of the last analysis.
The prediction values of the NIR chemical parameters are the result of the prediction AVERAGES of the
single analysis performed. In case of analysis not right performed, the AgriNIR™ system does not
consider them in the final average of the results.
In case a NIR chemical parameter following results :
n.a.
Parameter NOT AVAILABLE
This chemical parameter hasn’t been defined for the NIR family chosen in analysis.
Err.
Parameter with WRONG RESULT
The AgriNIR™ analyzer cannot predict this chemical parameter.
Please check the sample in the fodder box, the NIR family chosen or the NIR calibration.
Out.
Parameter OUT OF RANGE
This chemical parameter prediction value is out of the acceptability range.
In this picture you can see :
STARCH is not available for GRASS SILAGE
PROTEIN is out of range
ADF, NDF are not possible to predict
DRY MATTER
value of the sample :
calculated as 100% - Moisture.
MOISTURE (Humidity)
value of the
sample.
Chemical Parameters on DM
the chemical values expressed on the
dry matter :
calculated as Param
AsIs
/ DM%
Chemical Parameters on AsIs
values of the sample.