Use AgriNIR™ Analyzer Rev. A11
E
NG
L
IS
H
39
6.9
Case of incorrect analysis
In case of analysis (single or repeated) that are NOT correct, the AgriNIR™ system shows the following
error message :
If you press <ENTER>
the AgriNIR™
system proposes again to execute the analysis
on the same component just selected.
If you press <RESET>
the AgriNIR™
system quit the analysis execution phase on the
selected component.
TROUBLESHOOTING
1.
Verify if the sample is properly inserted inside the fodder box. Please see the
paragraph
Correct Use of the Fodder Box for the Samples to Analyze
at page 14.
2.
Verify if the fodder box is properly moved during analysis execution. Please see the
paragraph
Correct Use of the Fodder Box for the Samples to Analyze
at page 16.
3.
Verify if you have chosen the correct NIR family / component before start the analysis
on the sample.
4.
Verify the cleaning of the Fodder Box glass or the ReadNIR glass inside the NIR
group on the bottom side.
5.
Verify if the Fodder Box glass or the ReadNIR glass is broken, otherwise contact the
Customer Service for the spare parts.
6.
Verify if the lamp inside the NIR group is not on its correct position or its time life is
expiring (1 year); in second case, please contact the
Customer
Service for the spare part.
7.
Verify if the optic fiber terminal (visible inside the NIR group through the glass window
on the bottom side) is properly inserted on its brass fitting, otherwise if it's moved
please contact the
Customer Service.
8.
If it's none of the previous cases, please contact the
Customer
Service.