Use AgriNIR™ Analyzer Rev. A11
E
NG
L
IS
H
23
6.4
States & Messaging of NIR analyzer internal system
After turning on the AgriNIR™, before going on with the execution of the analysis it is better to wait
always at least
15 minute
before starting to allow :
•
the thermoregulation of the NIR optical sensor to reach the target temperature of 32°C;
•
the necessary warming up of the internal lamp of the Read NIR reading head;
•
one first calibration of the optical references of the internal NIR analysis system.
By the way, to help the user the AgriNIR™ system shows on the right top of the screen an icon referring
to the contemporary state of the NIR system:
AgriNIR
AgriNIR W
Description of the icon
WARM-UP
the internal NIR system is executing the initial warm-up phase to reach the
target temperature of 32°C and to warm-up of the internal lamp.
OPTICAL CALIBRATION
the internal NIR system is executing the calibration phase of the optical
references.
READY
the AgriNIR™ system is ready for the analysis execution.
RUNNING ANALYSIS
the AgriNIR™ system is executing the analysis.
CONNECTION MISSING
or
ERROR STATUS
the NIR group is not connected to the electronic unit or there is an issue the
internal NIR system.
ERROR OF NIR CALIBRATION
The system can’t performs the calibration.
After turning ON the machine
BEFORE STARTING
with the analysis executions, the AgriNIR™ system
executes automatically :
1.
the
WARM-UP phase
of the NIR group with status icon
, to allow :
the thermoregulation of the NIR optical sensor to reach the target temperature of 32°C;
the necessary warming up of the internal lamp of the Read NIR head;
2.
the
CALIBRATION phase
of the NIR group with status icon
, to allow the first calibration of
the optical references of the internal NIR analysis system;
3.
after these first two phases the AgriNIR™ analyzer is
READY
, with status icon
.
Although the WARM-UP takes less time and the CALIBRATION phase is
already done, it’s STRONGLY RECOMMENDED to wait always at least
15
minutes
before starting use the AgriNIR™ machine.
Please see
NOTES FOR THE USER point N° 1
at page 3.