![Dinamica Generale AgriNIR Operator'S Manual Download Page 31](http://html1.mh-extra.com/html/dinamica-generale/agrinir/agrinir_operators-manual_2505530031.webp)
Use AgriNIR™ Analyzer Rev. A11
E
NG
L
IS
H
29
4.
Press <ENTER>
to start the automatic testing procedure :
WARNING!
DO NOT move the fodder box with the
CHECK-CELL inside during testing
execution.
If you move the fodder box, the CHECK-CELL
cannot be in front of the internal optical reading
window (ReadNIR window, where is the lamp
light); consequently the CHECK-CELL cannot be
properly read.
Pressing <RESET>
testing execution is immediately aborted.
5.
After 40 seconds, the AgriNIR™ analyzer shows the optical test result :
The AgriNIR™ internal optics is OK!
Now you can proceed to execute analysis.
Press <ENTER>
or <RESET>
to
exit and continue.
6.
If the testing execution takes more than 40 seconds (2 minutes), the optics test could fail. In this case
the AgriNIR™ analyzer shows following message :
WARNING!
The AgriNIR™ internal optics is FAIL!
Consult the following troubleshooting procedure
to solve the issue.
Press <ENTER>
or <RESET>
to
exit and continue.
In this case you can save the optical spectra on
USB memory key pressing F-key
, then
send data to the
Customer
Service for checking.