Use AgriNIR™ Analyzer Rev. A11
26
TROUBLESHOOTING
1.
Please check if the communication cable from Electronic Unit (front control panel) is
connected to the NIR group inside the bottom box.
2.
Please check if the communication cable from Electronic Unit to NIR group is not
damaged.
3.
If is not the case 1 or 2, please contact the
Customer Service.
During functioning of the AgriNIR™ machine, if appears this following messages :
the thermoregulation of the internal NIR
optical sensor CANNOT maintain the target
temperature of 32°C.
Please IMMEDIATELY turn OFF the
AgriNIR™ machine and leave it off for at
least
30 minute
.
TROUBLESHOOTING
1.
Verify if the environmental temperature, where the AgriNIR™ machine is located, is
out of the
Working Temperature range
. Please see Technical Data at page 5.
2.
Remove completely the NIR group from the AgriNIR™ machine, then check if the
filter fan and the finned heat sink (visible on the back side where the shield is pierced)
are clean. Otherwise proceed to clean using a low flow of compressed air.
3.
Verify if the fan on the connector side of the NIR Group is working, otherwise contact
the
Customer Service for the spare part.
4.
If is not the case 1, 2 or 3, please contact the
Customer Service.