Use AgriNIR™ Analyzer Rev. A11
34
6.8
Analysis Execution with AgriNIR™ System
After turned ON the AgriNIR™ system and performed following phases :
•
WARM-UP (at least 15 minutes) and CALIBRATION
please see at page 19 for more details;
•
optics test with CHECK-CELL
please see at page 23 for more details;
•
identification of the sample (using SAMPLE ID
) and set the customer information
(CUSTOMER CODE
and CLIENT INFORMATION
);
it’s possible to prepare the sample inside the fodder box (see at page 14 for more details) and starts to
execute the analysis.
In the main screen it’s possible to select the component to analyze by pressing the keys
:
In case of
standard calibrations package, the portable analysis system AgriNIR™
can analyze till to a max of
7 NIR families
(feed components categories) as mentioned at page 13.
Moving the pointer you can choose among 7 NIR families / components (Corn Silage, Hay, High Moisture
Corn, Alfalfa Hay, Grass Silage, TMR, Soybean Flour) : every component that can be chosen in this list
represents a feed components category (the NIR family) which the sample to analyze can belong.
Select a different component / NIR families means select a different NIR calibration inside the AgriNIR™
analyzer.
Please take care to select the correct Component / NIR family before
starting the analysis on the forage sample.
The Component / NIR family selected represents the feed components
category which the sample to analyze belong.