Use AgriNIR™ Analyzer Rev. A11
36
The AgriNIR™ analyzer allows two analysis execution modes to satisfy following cases :
INSTRUCTIONS FOR PROPER SELECTION OF ANALYSIS MODE
Case of SAMPLES HOMOGENEOUS and LOW QUANTITY
Examples : High Moisture Corn, Soybean Flour
1. If the sample is already grinded not more than 2/3 mm, it's possible to put directly
the sample inside the fodder box.
2. Select
SINGLE ANALYSIS mode
to execute the analysis.
Case of SAMPLES NOT HOMOGENEOUS
Examples : Corn Silage, Hay, Alfalfa Hay, Grass Silage, TMR
1. Provide to shred the material to obtain pieces not longer than 2/3 cm, then put the
sample inside the fodder box.
2. Select
MULTIPLE ANALYSIS mode
, using
at least 3 analysis
.
3. Mix the sample inside the fodder box between a single analysis and the next.
Case of LARGE SAMPLE QUANTITY
Examples : bunker of Corn Silage, Grass Silage or TMR; bale of Hay or Alfalfa Hay
1. Collect enough samples that are representative of the entire bunker front, or the
entire bale, etc…
2. Provide to shred each samples to obtain pieces not longer than 2/3 cm, then put
one sample inside the fodder box.
3. Select
MULTIPLE ANALYSIS mode
, using
N° of analysis = N° of collected
samples
.
4. After executed a single analysis, take next the sample and put it into the fodder
box.
In case of
SINGLE ANALYSIS
execution mode (pressing <RESET>
) :
the AgriNIR™ system starts automatically with
the execution of a single analysis on the selected
component and therefore on the sample in the
fodder box.
WARNING!
Please start IMMEDIATELY to move the
fodder box as soon as you press
<RESET>
.
If you want to quit immediately the analysis, press<RESET>
during the analysis execution.