10-4 Built In Stage / Remote Stage Setup
XL™ Series
PC Interface View Built In Display Screen
D - Datum: 0.000
Datum [0.000 ]
This value is only used when the unit is set to the NOAA mode. This is a site specific offset
entered by the user and the value is transmitted as header information in the NOAA GOES data
format. This value is used to help decode the NOAA data.
PC Interface View Built In Display Screen
M - Measure
Not Available In Built In Menu
When using the PC menu, pressing the M key will cause the H-350XL™ to take a measurement
of the internal stage and temperature sensors and the sensor used for the remote stage, then
update the screen with the new values.
10.4 PtTmp Sensor Setup (H-350XL™ Model only)
Built in Menu Structure: Temperature Sensor Options
Sensor Input Setup->
PtTmp Sensor Setup->
PtTemp Units [Deg C]
The PC menu for this option is shown above.
PC Interface View Built In Display Screen
T - Temp Units: Deg C
PtTemp Units [Deg C]
This option is only available on the H-350 XL as it is the only XL series unit that has a built in
temperature sensor used for compensation of the pressure sensor. The temperature setup option
allows the user to select if temperature values from the built in temperature probe will be
displayed and recorded in degrees Celsius or degrees Fahrenheit. This only affects temperature
measurements made using the internal temperature sensor used for temperature compensation of
the pressure sensor. This is not a good sensor to use as an ambient air temperature sensor as the
temperature inside a gage house or enclosure may be several degrees different that the ambient
air temperature. If other temperature sensors are connected to the H-350XL™ via the generic
analog inputs or an SDI-12 input, they must be setup independently to produce the desired units.
The default temperature units selection for the internal temperature sensor is Celsius.
Summary of Contents for H-350XL
Page 1: ...XL Series Models H 350XL H 500XL H 510XL H 522 H 522 Plus Owner s Manual ...
Page 30: ...2 16 Hardware Options and Installation XL Series ...
Page 42: ...3 12 Using The Built In Keypad Display XL Series ...
Page 46: ...4 4 PC Menu Interface Operation XL Series ...
Page 58: ...5 12 Miscellaneous System Setup XL Series ...
Page 66: ...6 8 Serial Port Options XL Series ...
Page 74: ...7 8 System Configuration Files XL Series ...
Page 84: ...8 10 System Status Menus Options XL Series ...
Page 88: ...9 4 Scanning Options XL Series ...
Page 156: ...15 12 Logging Options XL Series ...
Page 198: ...16 42 Operation with the GOES Radio XL Series ...
Page 216: ...19 6 Alarm Call Out XL Series ...
Page 234: ...21 8 Operation with the H 355 Smart Gas System XL Series ...
Page 246: ...22 12 Functions XL Series ...