24-2 Maintenance/Trouble Shooting
XL™ Series
Artificially High Data:
1-
This could be an indication that the orifice in the water is restricted or plugged. Contact
the manufacturer of your gas purge system to find out how to purge the orifice. You may
2
want to change your supply gas from dry Nitrogen to Carbon Dioxide (CO ) for a short
period of time. This will create Carbonic Acid at the end of the orifice and will serve as a
cleaning agent to remove deposits.
2-
If you have entered a User Units Slope and/or Offset, either in our data recorder or in the
XL™, verify that they are calculated and entered correctly.
3-
Verify that the orifice line has not moved.
Artificially Low Data:
1-
This could be an indication of a leak somewhere in your system. Check all of your
plumbing connections with a soapy water solution or a commercial leak detection
solution such as SNOOP . Because the XL™ has laboratory quality accuracy, even the
®
smallest leak will show up as an error in your data.
2-
If you have entered a User Unit Slope and/or Offset, either in your data recorder or in the
XL™, verify they are calculated and entered correctly.
3-
Verify that the orifice line has not moved.
Wrong or Erratic Data:
1-
Check the sensor status window for possible errors. (See section 3.6)
2-
Check battery voltage and all connections.
GOES not transmitting:
1-
Check power to the GOES radio, do so under a load.
2-
See if the GOES time is correct, if it is stopped, it indicates power to the radio was lost,
stopping the clock. The radio will not transmit if the GOES clock is not running.
3-
Check all antenna wires.
4-
Check antenna alignment.
5-
Use a power meter to measure the power output of the radio, and the reflected power.
Summary of Contents for H-350XL
Page 1: ...XL Series Models H 350XL H 500XL H 510XL H 522 H 522 Plus Owner s Manual ...
Page 30: ...2 16 Hardware Options and Installation XL Series ...
Page 42: ...3 12 Using The Built In Keypad Display XL Series ...
Page 46: ...4 4 PC Menu Interface Operation XL Series ...
Page 58: ...5 12 Miscellaneous System Setup XL Series ...
Page 66: ...6 8 Serial Port Options XL Series ...
Page 74: ...7 8 System Configuration Files XL Series ...
Page 84: ...8 10 System Status Menus Options XL Series ...
Page 88: ...9 4 Scanning Options XL Series ...
Page 156: ...15 12 Logging Options XL Series ...
Page 198: ...16 42 Operation with the GOES Radio XL Series ...
Page 216: ...19 6 Alarm Call Out XL Series ...
Page 234: ...21 8 Operation with the H 355 Smart Gas System XL Series ...
Page 246: ...22 12 Functions XL Series ...