12-8 Digital I/O
XL™ Series
PC Interface View Built In Display Screen
S - Encoder Slope = 0.010
Encodr Slp[0.01000]
This option screen allows the user to enter a slope (or multiplier) value that is applied to each
count in order to calculate the final quadrature input value. The default slope is 0.01. Most shaft
encoder’s produce 100 counts per revolution, so a slope of 0.01 will equate to 1.00 for one
revolution. When using a pulley with a 1 foot circumference, each count is 0.01 foot of change
and one revolution is one foot of change.
PC Interface View Built In Display Screen
Not Available Using PC Menu
Test Quad Out?
This option allows the user to test the simulated encoder output mode. Pressing the Enter key on
this screen will prompt the user to
Press Up or Down
. Pressing the up key will cause the XL™
to simulate 100 counts up, whereas pressing the down key will simulate 100 counts down.
Normally the user will first count up 100 counts and then do the test again by counting down 100
counts so the net change is 0.0 but looking on a chart drive a short vertical line would be visible.
Summary of Contents for H-350XL
Page 1: ...XL Series Models H 350XL H 500XL H 510XL H 522 H 522 Plus Owner s Manual ...
Page 30: ...2 16 Hardware Options and Installation XL Series ...
Page 42: ...3 12 Using The Built In Keypad Display XL Series ...
Page 46: ...4 4 PC Menu Interface Operation XL Series ...
Page 58: ...5 12 Miscellaneous System Setup XL Series ...
Page 66: ...6 8 Serial Port Options XL Series ...
Page 74: ...7 8 System Configuration Files XL Series ...
Page 84: ...8 10 System Status Menus Options XL Series ...
Page 88: ...9 4 Scanning Options XL Series ...
Page 156: ...15 12 Logging Options XL Series ...
Page 198: ...16 42 Operation with the GOES Radio XL Series ...
Page 216: ...19 6 Alarm Call Out XL Series ...
Page 234: ...21 8 Operation with the H 355 Smart Gas System XL Series ...
Page 246: ...22 12 Functions XL Series ...