HD31
- 114 -
V1.6
RECORD function (statistics):
Command
Response
Description
RCDCHN:cc
(cc = 01…36)
RCDCHN:cc;n|
Indicates whether the measurement with
sequential number cc on the display of the
instrument is selected for RECORD function:
not selected if n=0
selected if n=1
RCDCHN:cc;n
(cc = 01…36; n=0,1)
RCDCHN:cc;n; ok|
Selects/deselects the measurement with
sequential number cc on the display of the
instrument for the RECORD function:
deselects if n=0
selects if n=1
RCDMAN:cc
(cc = 01…36)
RCDMAN:cc; ok|
Manual acquisition, for the RECORD func-
tion, of a sample of the measurement with
sequential number cc on the display of the
instrument.
If measurements for the RECORD function
have not been selected, the command
enables the RECORD function for all the
measurements.
RCDSTATUS:cc
(cc = 01…36)
RCDSTATUS:cc;n;m|
Indicates the RECORD function status for
the measurement with sequential number
cc on the display of the instrument:
function disabled if n=0
function enabled and automatic acqui-
sition enabled if n=1
function enabled and automatic acqui-
sition disabled if n=2
current value displayed if m=0
minimum value displayed if m=1
average value displayed if m=2
maximum value displayed if m=3
RCDSTATUS:cc;n;m
(cc = 01…36;
n=0,1,2; m=0,1,2,3)
RCDSTATUS:cc;n;m; ok| Sets the RECORD function for the meas-
urement with sequential number cc on the
display of the instrument:
function disabled if n=0
function enabled and automatic acqui-
sition enabled if n=1
function enabled and automatic acqui-
sition disabled if n=2
current value displayed if m=0
minimum value displayed if m=1
average value displayed if m=2
maximum value displayed if m=3
RCDSYNCH:cc
(cc = 01…36)
RCDSYNCH:cc; ok|
Reset of statistics