HD31
- 90 -
V1.6
13
CALIBRATION
The probes are supplied factory-calibrated and usually don’t require additional user interven-
tion. Nevertheless a new calibration can be performed.
Note
: it is essential to know and respect the physical phenomena measurements are based on
in order to achieve correct calibration: it is recommended to perform new calibrations only if
you have the adequate technical expertise and to carefully follow the procedures described in
this manual.
To access probe calibration:
1.
Press
SELECT
to display the
MENU
command in the command bar,
then select the command by press-
ing the
F2
key.
2.
Select the
CALIBRATION
item using
S
and
T
, then press
the
F3
key (
OK
command) to confirm.
Note
: the CALIBRATION item is not displayed if the instru-
ment is in protected mode. See paragraph 8.2 to change the
protection status.
3.
The list of the connected probes is displayed. Each probe is
identified by its own serial number and by the type of meas-
urement. Select the probe to be calibrated using
S
and
T
,
then press the
F3
key (
OK
command) to confirm.
4.
The displayed screen depends on the type of probe selected. According to the type of probe,
continue as indicated in the paragraphs below.
13.1
C
ALIBRATION OF TEMPERATURE
-
ONLY PROBES WITH
PRT
SENSOR
Selecting a Pt100 or Pt1000 temperature-only probe for calibration displays four possibilities:
•
NEW TUNING
: calibrates the probe in one, two or three points. The possible calibration
points are 0 °C, one point ranging from 95 °C to 105 °C and one point ranging from
150 °C to 400 °C. The probe can be calibrated either in one point or in two points: for
the non-calibrated points, the instrument will use the values stored in the previous user-