HD31
- 73 -
V1.6
5.
For the automatic temperature compensation, select, using
S
and
T
, the probe providing the temperature measurement
(PROBE A, PROBE B or PROBE C), then press the
F3
key (
OK
command) to confirm.
Note
: the current setting appears in white color, while the
other available settings are yellow to highlight that they are
not the current setting.
For manual temperature compensation, select, using
S
and
T
, the
MANUAL ENTRY
item, then press the
F3
key (
OK
command) to confirm.
6.
For manual compensation, confirming the MANUAL ENTRY op-
tion displays the current setting of the temperature value.
Select the
UNIT
command by pressing the
F1
key to change
the measurement unit (°C, °F or K).
7.
To set the temperature value, select the
EDIT
command by pressing the
F3
key.
8.
Set the first digit using
S
and
T
, then select the "-->" com-
mand (
F2
key) to move to the next digits (the currently se-
lected digit flashes). To return to a previous digit, select the
"<--" command (
F1
key). After setting all the digits, press
OK
(
F3
key) to confirm the value.
The value must be set to two decimal digits after the comma
and must be in the range between -30 °C and +120 °C (or
the values equivalent in other units of measurement).
Note
: the value set in °F is converted internally in °C, there-
fore it is possible to verify a deviation of ± 0.1 °F between
the value set by the user and the actual stored value, due to
the rounding of the conversion result.
9.
Repeatedly press the
ESC
button to exit the menu.