HD31
- 128 -
V1.6
PP473 S0
SICRAM probe for the measurement of relative pressure with respect to the at-
mosphere or differential pressure in the range ± 250 Pa.
The probe uses a silicon piezoresistive sensor with high accuracy and temperature
compensation, which has excellent linearity, repeatability and stability over the
time.
A special
auto-zero
circuit periodically equalizes the differential pressure at the
sensor input and corrects the offset; this feature makes the probe insensitive to
the mounting position and compensates the sensor aging and the deviation of the
zero with temperature changes, virtually eliminating maintenance operations.
A typical application of the probe is clean rooms monitoring.
Measuring range
± 250 Pa (± 2.5 mbar)
Maxuimum overpressure
50 kPa
Resolution 0.1
Pa
Accuracy @ 25 °C
± (0.2 Pa + 1.5% of the measure)
Accuracy @ 0…50 °C
± (0.2 Pa + 3% of the measure)
Operating temperature
-10…+60 °C
Response time
0.125 s
Long-term stability
± 0.5% f.s. nominal (1000 h @ 25 °C)
Fluid in contact with
Non-corrosive dry gas or air
the membrane
Connection
Ø 5 mm Tube
PP473 S1...PP473 S8
SICRAM probes for the measurement of the differential pressure
Measuring range
S1
=f.s. 10 mbar
S2
=f.s. 20 mbar
S3
=f.s. 50 mbar
S4
=f.s. 100 mbar
S5
=f.s. 200 mbar
S6
=f.s. 500 mbar
S7
=f.s. 1 bar
S8
=f.s. 2 bar
Max overpressure
S1,S2,S3
=200 mbar
S4
=300 mbar
S5,S6
=1 bar
S7
=3 bar
S8
=6 bar
Accuracy @ 25 °C
S1,S2,S3
=0.5% f.s.
S4
=0.25% f.s.
S5,S6,S7,S8
=0.15% f.s.
Operating temperature
-10...+60 °C
Fluid in contact with
Air or non-corrosive dry gases
the membrane
Connection
Pipe Ø 5 mm