HD31
- 77 -
V1.6
4.
Select the
AIR FLOW AREA
item using
S
and
T
, then press
the
F3
key (
OK
command) to confirm.
5.
The section current setting appears. Select the
UNIT
com-
mand by pressing the
F1
key to change the measurement
unit (cm
2
or in
2
).
6.
To set the section value, select the
EDIT
command by pressing the
F3
key.
7.
Set the first digit using
S
and
T
, then select the "-->" com-
mand (
F2
key) to move to the next digits (the currently se-
lected digit flashes). To return to a previous digit, select the
"<--" command (
F1
key). After setting all the digits, press
OK
(
F3
key) to confirm the value.
If cm
2
is used as measurement unit, the set value must be
integer and range from 10 to 100000 cm
2
. If in
2
is used as
measurement unit, the set value must have a decimal digit
after the comma and range from 1.5 to 15500.0 in
2
.
Note
: the value set to in
2
is internally converted into cm
2
,
therefore a deviation of ± 0.1 in
2
between the value set by
the user and the value actually stored may occur, due to the
rounding up of the conversion result.
8.
Repeatedly press
ESC
to exit the menu.