HD31
- 71 -
V1.6
4.
Select the probe connected to A, B or C input
using the
PROBE
command (
F1
key).
5.
Use
S
keys
T
to change the setting of the type of calibration for the selected probe. The
available settings are:
•
FACTORY
: factory calibration is used, even if an user calibration is available.
•
USER
: user calibration is used, if available, even if it has been performed with an in-
strument different from the one the probe is connected to. If the user calibration is not
available, factory calibration is used.
•
AUTO
: user calibration is used, if available, only if it has been performed with the in-
strument the probe is connected to.
Note
: if a user calibration is not available for the selected probe, the default setting is AUTO
and cannot be changed.
Note
: the type of calibration cannot be changed if the instrument is in secure mode. See pa-
ragraph 8.2 to change the protection status.
Note
: the current setting appears in white, while the other available settings are shown in
yellow to highlight that they are not the current setting.
6.
Press
F3
(
OK
command) to confirm the setting for the selected probe. A message appears
asking to confirm the change of calibration type: select
YES
(
F3
key) to continue (or press F1
to select NO and exit).
7.
Repeatedly press
ESC
to exit the menu.
Serial number of
selected probe
Selected
input