![DeltaOHM HD31 Operating Manual Download Page 110](http://html1.mh-extra.com/html/deltaohm/hd31/hd31_operating-manual_2479063110.webp)
HD31
- 110 -
V1.6
14.1
S
TORED DATA TRANSFER TO
PC
Each time a logging session is launched, the instrument opens a new file, in CSV format,
where the measurements acquired within the end of the session are stored. The file is created
in a folder identified by the logging start time (e.g. "R_164905" if the logging operation has
been started at 16:49:05), in turn contained in a folder identified by the logging start date
(e.g. "D_150513" if the logging operation has been started on 13/05/2015).
Fig. 14.1.1: structure of the SD memory card
The report in PDF format is also contained in the same folder of the CSV file of the logging ses-
sion, if generated.
The stored data can be transferred to PC in the following ways:
•
Removing the SD memory card from the instrument and inserting it in a card reader
connected to your PC.
•
Connecting the USB port of the instrument to your PC through the
CP31
cable and with
the help of DeltaLog9 application software. Additionally, the use of DeltaLog9 software al-
lows the analysis of the downloaded data through the processing functions included in
the software. See the software instructions for data download through DeltaLog9 soft-
ware.
•
Connecting the USB port of the instrument to your PC through the
CP31
cable and set-
ting the instrument to "card reader" mode; thanks to this mode, the SD card inserted in
the instrument is read by the PC as a disc unit (therefore, it allows data files to be cop-
ied using Windows
®
Explorer).
To set the instrument to "card reader" mode:
1.
Press
SELECT
to display the
USB
in the command bar, then select
the command by pressing the
F3
key.
Note
: USB command is disabled if
logging is in progress or the in-
strument is not connected to the
PC.
2.
A message appears reminding that the “card reader" mode requires the measurement proc-
ess to be stopped (and consequently the automatic RECORD function as well, if enabled).
Select
YES
(
F3
key) to continue (or Press F1 to select NO and exit).
3.
In “card reader" mode, the display shows the image of a SD card. Select again the
USB
command by pressing the
F3
key to return to the measurement mode; a message will ap-
pear asking to confirm exit from the “card reader" mode, select
YES
(
F3
key) to exit.
Note
: the exit from the “card reader" mode is automatic if you disconnect the USB cable.