Expert
Logger
The example means: 0 mA corresponds to 0 l/h (litres per hour), 20 mA
corresponds to 1000 l/h.
Sensor type
Select the required fixed measuring range in
Sensor type
or in the
uni-
polar
/
bipolar
settings in
Measuring range
.
unipolar
: Measuring only positive signals.
bipolar
: Measuring positive and/or negative signals.
Shunt
: Enter the load resistance being used.
Delphin
offers various load
resistances for the current measurement. Contact us to obtain a suitable
resistance.
Current measurement is not possible without load res-
istance.
Scaling
Enter two points from the sensor's characteristic curve, if possible, the
lower and upper vertexes. For a linear characteristic, measurement values
are then converted into the unit of measurement (scaling) from the
sensor. For non-linear characteristics use
Sensor compensation
. See
tab. The specified range (in the example 0 l/h to
1000 l/h) is used as the default for the
ProfiSignal
graphic display area.
The channel receives the status "scale range overrun" or "scale range
underrun" when not within the range. Channel status values are available
in the table in
.
4.10.1.3
Resistance sensor type
95
Delphin Technology AG
Januar 2018
Expert
Logger
Manual