Expert
Logger
Level triggering
: When a level moves from an active to inactive state, the
selected function is applied to the values accumulated during the active
level period. The result is output and the accumulated values then deleted.
Generate intermediate values
: When enabled, the results between the
trigger events are immediately output after calculation (with each new
source value).
Refer also to
Definitions
l
The minimum is the smallest value within the data series.
l
The maximum is the greatest value within the data series.
l
The variance is the mean square error relative to the mean.
l
The number of measurement values are values in the block interval.
l
The standard deviation is the square root of the variance (not the average
deviation!).
l
The True RMS is the root from the sum of the squares of the values by the
number of values.
l
The arithmetic mean, also called the average, is the sum of all values of a
data series divided by the number of values.
l
The geometric mean is the Nth root from the product of n values.
l
The harmonic mean is the reciprocal from an average of reciprocals.
4.11.21
Status filter
The function allows certain status messages to be suppressed and another
output to be controlled. You can pass on all status messages to a digitial
output to be informed of problems, but relay only certain messages to a
second digitial output, such as the actions of switching off machines etc.
Refer to
.
Available status functions and their meaning
Selection
Description
Wire break
Wire break in monitored channel.
Configuration
error
Error in monitored channel configuration.
Invalid value
Invalid value in monitored channel.
Measuring
range over-
run
The measured value is outside the (positive) measuring
range of the A/D converter, that is, it is too high in the pos-
itive range.
Measuring
range under-
The measured value is outside the (negative) measuring
range of the A/D converter, that is, it is too high in the neg-
163
Delphin Technology AG
Januar 2018
Expert
Logger
Manual