DT9100 B
Sequence of operations
11. If you want to test synchronisation input, press ▲
button to change
mnemonic to
mnemonic.
(Its meaning: synchronisation input test)
12. Press
ENTER
button.
13. You will see on the display:
•
when the synchronisation input is in non-active state
(terminals 7 and 8 are, R > 200 kΩ)
or you will see on the display:
•
when the synchronisation input is in active state (terminals
7 and 8 are closed, R < 1 kΩ)
14. Press
MENU
button to exit this menu item:
you will see on the display the
mnemonic. The
mnemonic will be blinking.
15. Press the ▲ button to change the
mnemonic to
the
mnemonic. (Its meaning: Temperature
measurement)
16. Press
ENTER
button.
17. You will see on the display the internal temperature of
the instrument in °C.
18. Press
MENU
button to exit this menu item: you will see
on the display: the
mnemonic. The mnemonic will
bee blinking.
Exit from the menu item
1. Press the
MENU
button, to exit from the Tests menu
item, and you see this:
.
(2. If you want to return to the tests, go on from
point 3 of
the Sequence of operations
.)
(3. If you don’t want to exit the menu because you want to
perform further settings, you can select the desired menu
items by pressing the ▲ button.)
Exit from the menu
1. Press the
MENU
button. First the mnemonic
(
Save
), then the mnemonic
(
Exit
) are shown on the
display. With this the storing of the settings is completed.
The instrument has exited the menu and goes on with the
measurement.
20180905-V1
43
Summary of Contents for DT9100 B
Page 1: ...DT9100 B Intrinsically Safe Digital Clock Operating Instructions...
Page 47: ...DT9100 B 10 2 Application example 20180905 V1 47...
Page 56: ...DT9100 B 10 8 ATEX Certification 56 20180905 V120180905...
Page 57: ...DT9100 B 20180905 V1 57...
Page 58: ...DT9100 B 58 20180905 V120180905...
Page 59: ...DT9100 B 20180905 V1 59...
Page 60: ...DT9100 B 60 20180905 V120180905...
Page 61: ...DT9100 B 20180905 V1 61...
Page 62: ...DT9100 B 62 20180905 V120180905...
Page 63: ......