Product Reference Guide
6-9
between 5.995 and 6.005. Remove the weights and verify that the
display reads 0.00.
5. You have completed the Increasing-Load Test (Phase 1).
Shift Test (6 Pounds)
The Shift Test checks to ensure that items placed anywhere on the weigh-
ing surface of the scanner/scale are weighed properly. Refer to
Figure 6-2
when performing this test.
1. Place and remove, in succession, a six-pound load on the center of
each of the four quadrants: 1, 2, 3, and 4, (as shown in
Figure 6-2
)
of the scanner/scale’s weigh platter. Verify that a reading of
between 5.995 and 6.005 lb. registers, under load, in each quad-
rant tested. Check that the display returns to 0.00 between each
load.
2. After verifying the accuracy of each quadrant of the weighing sur-
face, remove all weight from the scale. This completes the Shift
Test.
Figure 6-2. Shift Test
1
2
3
4
1
2
4
3
(Models vary)
Summary of Contents for Magellan 8500Xt
Page 1: ...MagellanTM 8500Xt Product Reference Guide...
Page 14: ...12 MagellanTM 8500Xt Scanner NOTES...
Page 70: ...2 40 MagellanTM 8500Xt Scanner NOTES...
Page 88: ...3 18 MagellanTM 8500Xt Scanner Figure 3 6 Removing Replacing the Gasketed DLC Window a b c d...
Page 93: ...Product Reference Guide 4 5 Figure 4 1 7 Segment LED Display 7 Segment Display Models vary...
Page 102: ...4 14 MagellanTM 8500Xt Scanner NOTES...
Page 138: ...6 20 MagellanTM 8500Xt Scanner NOTES...
Page 416: ...7 278 MagellanTM 8500Xt Scanner NOTES...
Page 428: ...A 12 MagellanTM 8500Xt Scanner...
Page 448: ...D 2 MagellanTM 8500Xt Scanner NOTES...
Page 462: ...E 14 MagellanTM 8500Xt Scanner NOTES...
Page 477: ...Product Reference Guide F 15 NOTES...