Datacolor CONDITIONER™ (MB2) 06/2017
page 13/126
1.8
Measures to prevent accidents
The operator of the cabinet must observe the following rule: “Betreiben von Arbeitsmitteln. Betreiben von
Kälteanlagen, Wärmepumpen und Kühleinrichtungen“ (Operation of work equipment. Operation of refrig-
eration systems, heat pumps and refrigeration equipment) (GUV-R 500 chap. 2.35) (for Germany).
The manufacturer took the following measures to prevent ignition and explosions:
•
Indications on the type plate
See operating manual chap. 1.4.
•
Operating manual
An operating manual is available for each cabinet.
•
Overtemperature monitoring
The cabinet is equipped with a temperature display, which can be read from outside.
The cabinet is equipped with an additional safety controller (temperature safety device class 3.1 acc.
to DIN 12880:2007). Visual and audible (buzzer) signals indicate temperature exceeding.
•
Safety, measurement, and control equipment
The safety, measuring, and control equipment is easily accessible.
•
Electrostatic charge
The interior parts are grounded.
•
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electri-
cal equipment (e.g. electric motors, power cables, solenoids). The machine has no permanent mag-
nets. If persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of
field source to implant) of 30 cm, an influence of these implants can be excluded with high probability.
•
Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
•
Floors
See operating manual chap. 3.4 for correct installation
•
Cleaning
See operating manual chap. 20.3.
•
Examinations
The chamber has been inspected by the “Deutsche Gesetzliche Unfallversicherung e.V. (DGUV)
(German Social Accident Insurance (DGUV)” (German Social Accident Insurance (DGUV), Testing
and Certification Body
for
Foodstuffs and Packaging Industry in DGUV Test).