Datacolor CONDITIONER™ (MB2) 06/2017
page 23/126
4.2
Wastewater connection
Fasten the wastewater hose to the wastewater connection “OUT” (7) on the rear of the cabinet (olive
∅
14 mm). Observe the following points:
•
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against at least 95 °C /
203 °F
.
•
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
•
Protect the cabinet end of the drainage hose with one of the supplied hose clamps.
•
Reliably prevent sucking back of wastewater. The end of the wastewater hose must not be immersed
in liquids. This can be ensured e.g., by free discharge.
Waste water is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous waste water flow.
Protect the waste water supply with the supplied hose clamps.
4.3
Freshwater supply
Connect the waste water pipe before connecting the cabinet to a freshwater pipe or filling the
freshwater can (option, chap. 4.5).
You can supply the cabinet with freshwater via a water pipe or by manually filling a freshwater can (op-
tion, chap. 4.5).
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
CAUTION
Calcification of the humidifying system.
Damage to the cabinet.
Operate the cabinet with deionized (demineralized) water only.
Types of suitable water quality:
•
Deionized water from a water treatment installation already existing at the customer's site. Conductivi-
ty from 1 µS /cm up to a maximum of 20 µS/cm. (Water, which is in equilibrium with the CO
2
in the air,
and has a conductivity below 1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH.)
•
Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable sys-
tem). A reusable measuring equipment to assess the water quality is included (chap. 4.4).
The manufacturer is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by the manufacturer.
The warranty becomes void in the event of use of water of deviating quality.