Datacolor CONDITIONER™ (MB2) 06/2017
page 47/126
7.1
Set-point entry for temperature, humidity, light intensity, and fan speed
through the “Setpoints” menu
Press the
Setpoint setting
icon to access the “Setpoint” setting menu from Normal display.
“Setpoints” menu (cabinet without lights)
“Setpoints” menu (cabinet with lights)
Select “Fixed value operation setpoints” to access the individual parameters.
•
Select the field “Temperature” and enter the desired temperature setpoint.
Cabinet without lights: Setting range: -5 °C up to 70 °C.
Cabinet with lights: Setting range: -5 °C up to 50 °C.
Confirm entry with
Confirm
icon.
•
Select the field “Humidity” and enter the desired humidity setpoint.
Cabinet without lights: Setting range: 0% r.H. up to 80% r.H.
Cabinet with lights: Setting range: 0% r.H. up to 70% r.H.
Confirm entry with
Confirm
icon.
•
Select the field “Fan” and enter the desired fan speed setpoint.
Setting range: 40% up to 100% fan speed.
Confirm entry with
Confirm
icon.
•
Cabinet with lights: Select the field “Light” and enter the desired light intensity setpoint.
Setting range: 0% up to 100% light intensity. With setting 0% the light is off.
Confirm entry with
Confirm
icon.
When entering a value outside the setting range, the message: “Value outside of limits! (Min:
xxx, Max: xxx)” appears (xxx is a wildcard for the limits of the respective parameter). Press the
Confirm
icon and repeat the entry with a correct value.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu,
or
press
the
Close
icon to exit the menu without taking over the entries.