Datacolor CONDITIONER™ (MB2) 06/2017
page 122/126
Cabinet
without lights
Cabinet
with lights
Illumination data per light cassette
Light intensity, setting 100%
Lux
---
4500
Environment-specific data
Noise level (mean value)
dB (A)
52
52
Energy consumption at 21 °C /
69.8 °F
and 65% r.H.
without illumination
Wh/h
≤ 500
≤ 500
Energy consumption at 21 °C /
69.8 °F
and 65% r.H.
with illumination
Wh/h
---
≤ 580
Filling weight of refrigerant R 134A (GWP 1300)
kg
0,180
0,180
*
Cabinet with lights: All climatic data established with illumination
Illumination data:
Average value, measured at +22 °C +/- 3 °C /
71.6 °F +/- 5.4 °F
with a spherical sen-
sor (+/-10%) by 12 cm /
4.7 in
below the light cassette.
All technical data is specified for unloaded cabinets with standard equipment at an ambient temperature
of +22 °C +/- 3°C /
71.6 °F +/- 5.4 °F
and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER factory standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for cabinets produced in series. We reserve the right to
change technical specifications at any time.
If the cabinet is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifi-
cation and may affect the humidity accuracy.
23.5 Equipment
To operate the cabinet, use only original Datacolor accessories or accessories / components
from third-party suppliers authorized by Datacolor. The user is responsible for any risk arising
from using unauthorized accessories.
Regular equipment
Microprocessor display program controller with 2-channel technology for temperature and humidity
Ethernet interface for computer communication
Electronic safety controller (temperature safety device) class 3.1 acc. to DIN 12880:2007
Inner glass door with gasket
DCT™ refrigerating system with refrigerant R134a
Microprocessor controlled humidifying and dehumidifying system *) (humidity range, see diagram)
2 racks, stainless steel
Access port 30 mm with silicone plug
Illumination system: “Bio Vital” fluorescent tubes, 2 light cassettes with 2 fluorescent tubes each
*) A water supply (1 to 10 bar) is necessary for the installation of the humidifying and de-humidifying
system (chap. 4.3). If no suitable house water connection is available, you can manually supply water
by filling a freshwater can (option, chap. 4.5). Furthermore, a water drain in a max. distance of 3 me-
ters /
9.8 ft.
and a max. height of 1 meter /
3.3 ft.
is required (chap. 4.2).