Datacolor CONDITIONER™ (MB2) 06/2017
page 32/126
(A)
Carrying handle
(B)
Pilot lamp (green)
(C)
Power switch
(D)
Connection socket for cabinet
(E)
Power cable
Figure 15: Voltage changer (front)
To establish the electrical connection of the cabinet with the voltage changer, proceed in the following
order:
1. Connect the power cable of the cabinet to the connection socket (D) of the voltage changer
2. Establish the power connection of the voltage changer. The socket must provide a protective conduc-
tor.
3. Turn on the voltage changer at the power switch (C) (position “I”). The green pilot lamp (B) lights up.
4. Turn on the cabinet with the main power switch (1) in the lateral control panel
Position “0” = off
Position “I” = on
Figure 16: Power switch of the voltage changer
Dimensions of the voltage changer
Width
mm
255
Depth (without door handles)
mm
360
Depth (incl. cable and door handles)
mm
450
Height
mm
300
Length of the connection cable to wall socket
mm
172
Lateral wall clearance of the cabinet
to set up the voltage changer (minimum)
mm
400
Electrical connection data of the voltage changer
Input side
V
115
A
26,9
Output side (to the cabinet)
V
214
A
13,0
Power frequency
Hz
50 / 60
(A)
(B)
(C)
(D)
(E)