11. Notification and alarm functions
11.1 Notification and alarm messages overview
11.1.1 Notifications
Notifications are indicated by
information icons
displayed in the screen header in Normal display
An information icon serves as an indication of a certain condition.
If this condition persists, in some cases an alarm will be triggered after a fix or configurable interval. As
long as the condition persists, the information icon therefore continues to be displayed also in state of
alarm. If during alarm the conditions ends, e.g., if during a tolerance range alarm the actual value returns
to within the tolerance range, the information icon disappears, whereas the alarm will continue until man-
ual acknowledgement.
Press the flash icon next to the information icon to access the corresponding text information.
Normal display showing the text information.
The currently valid information texts are high-
lighted in black (example:
Idle mode
)
Condition
Information
icon
Text information
Start after
condition occurred
The controller is in Idle mode (chap. 5.4).
“Idle mode”
immediately
The current actual temperature value is out-
side the tolerance range (chap. 11.4)
“Temperature
range”
immediately
The current actual humidity value is outside
the tolerance range (chap. 11.4)
“Humidity range”
immediately
The humidification / dehumidification system
is turned off (via operation line and/or by
setting “Control on/off”)
or
Temperature setpoint below 0 °C or above
95 °C
“Humidity off”
immediately
Cabinet door open
“Door open”
immediately
Notifications are not shown in the event list.