Datacolor CONDITIONER™ (MB2) 06/2017
page 61/126
Section view indicating the operation lines.
Activated operation line: switching status “1” (On)
Deactivated operation line: switching status “0” (Off)
The operation lines count from right to left.
Example:
Activated operation line “Humidity off” = 000000000000000
1
Deactivated operation line “Humidity off” = 000000000000000
0
9.7.4 Setpoint entry
•
Select the field “Temperature” and enter the desired temperature setpoint.
Cabinet without lights: Setting range: -5 °C up to 70 °C.
Cabinet with lights: Setting range: -5 °C up to 50 °C.
Confirm entry with
Confirm
icon. The controller returns to the section view.
•
Select the field “Humidity” and enter the desired humidity setpoint.
Cabinet without lights: Setting range: 0% r.H. up to 80% r.H.
Cabinet with lights: Setting range: 0% r.H. up to 70% r.H.
Confirm entry with
Confirm
icon. The controller returns to the section view.
•
Select the field “Fan” and enter the desired fan speed setpoint.
Setting range: 40% up to 100% fan speed.
Confirm entry with
Confirm
icon. The controller returns to the section view.
•
Cabinet with lights: Select the field “Light” and enter the desired light intensity setpoint.
Setting range: 0% up to 100% light intensity. With setting 0% the light is off.
Confirm entry with
Confirm
icon. The controller returns to the section view.
9.7.5 Tolerance range
You can specify a temperature and humidity program tolerance range for each program section with dif-
ferent values for the tolerance minimum and maximum. When the actual value exceeds the given thresh-
old, the program is interrupted. This is indicated on the display (see below). When the actual temperature
is situated again within the entered tolerance limits, the program automatically continues. Therefore, the
duration of the program may be extended due to the programming of tolerances.