Pulse CO-Oximeter settings:
OPTION
FACTORY DEFAULT
SETTING
CONFIGURABLE SETTING
DISPLAY VIEW
Set to Pleth + SigIQ
Pleth + SigIQ, Pleth Only,
AVERAGING TIME
Set to 8
2, 4, 8, 10, 12, 14, or 16 seconds
FASTSAT
Set to No
Yes/No
DATE FORMAT
Set to mm/dd/yy
mm/dd/yy and dd/mm/yy
TIME FORMAT
Set to 12 hr.
12 and 24 hour
PULSE BEEP VOLUME
Set to Level HIGH
Level LOW, MED, HIGH
TREND DISPLAY
Set to %SpO2 + BPM
See Section 4, Display for all
settings.
TREND PERIOD
Set to Autoscale.
See Section 4, Trend Setup for all
settings.
LOW SpO2 ALARM LIMIT
Set to 90
SpMet HIGH/LOW ALARM LIMIT
Set to High: 3.0/Low: ---
SpCO HIGH/LOW ALARM LIMIT
Set to High: 10/Low: ---
SMART TONE
Set to No
Yes/No
PULSE RATE HIGH/LOW
ALARM LIMIT
Set to High: 140/Low: 50
BPM
SpMet HIGH/LOW ALARM LIMIT
Set to High: 3.0/Low: ---
SpCO HIGH/LOW ALARM LIMIT
Set to High: 10/Low: ---
ALARM SILENCE
Set to 120 seconds
ALARM DELAY
Set to 5 seconds