DM DOC 040 Rev 1.0
LNOP® REUSABLE SENSORS
(LNOP sensors must be used in conjunction with Red PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low
Perfusion
Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse
Rat
LNOP DCI
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP DCIP 10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP YI
> 1 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
N/A
N/A
LNOP TC-I
> 30 kg
± 3.5%
N/A
± 3 bpm
N/A
± 3.5%
± 3 bpm
LNOP DC-
195
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP TF-I
> 30 kg
± 2%
N/A
± 3 bpm
N/A
± 2%
± 3 bpm
NOTE:
The LNOP TF-I and TC-I sensors were not validated under motion condition
LNOP® ADHESIVE SENSORS
(LNOP sensors must be used in conjunction with Red PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOP Adt
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Adtx
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Pdt
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Pdtx
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Neo
< 10 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP NeoPt < 1 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP Neo-L < 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
> 40 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP
NeoPt-L
< 1 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP Inf-L
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOPvTM ADHESIVE SENSORS
(LNOPv sensors must be used in conjunction with Red PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOPv In
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOPv Ne
< 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOPv Ad
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
© 2013 DanMedical Ltd
67