DM DOC 040 Rev 1.0
Set up
Remove the instrument from the shipping carton and
examine it for signs of damage.
Before use, the D-MAS battery needs to be fully charged:
Insert battery
Connect the AC adapter to the D-MAS power
connector.
Connect the AC power adapter into an AC power
source.
Verify that the batteries are charging.
D-MAS battery charging LED indicators illuminated amber
during charging and green when batteries are fully
charged.
Shut down procedure
Click start/shut down. Alternatively, use the power on/off
control.
Initial installation
Place D-MAS on a stable hard flat surface near the patient. Always place D-MAS on a dry surface
and maintain a minimum of 3 cm (1 inch) free space around the unit. Make sure that the D-MAS
loudspeaker is not covered to avoid a muffled alarm sound.
Configure the unit for your regional power line frequency (50 or 60 hz).
CAUTION:
THE UNIT MUST BE CONFIGURED TO MATCH YOUR LOCAL
POWER LINE FREQUENCY TO ALLOW FOR THE CANCELLATION
OF NOISE INTRODUCED BY FLUORESCENT LIGHTS AND OTHER
SOURCES.