![DanMedical D-MAS HyperSat 2540 Instructions For Use Manual Download Page 95](http://html1.mh-extra.com/html/danmedical/d-mas-hypersat-2540/d-mas-hypersat-2540_instructions-for-use-manual_3416622095.webp)
DM DOC 040 Rev 1.0
Pulse CO-Oximeter measurement range
Oxygen Saturation (%SpO2):
0
-100%
Carboxyhemoglobin Saturation (%SpCO):
0
- 99%
Methemoglobin Saturation (%SpMet):
0
- 99.9%
Pleth Variability Index
0 - 99%
Pulse Rate:
25 - 240 (bpm)
Perfusion Index:
0.02% - 20%
ACCURACY
Saturation
70% to 100%
No Motion1
Adults, Pediatrics
±2 digits
Neonate
2
±3 digits
Motion3
Adults, Pediatrics
±3 digits
Neonate
±3 digits
Low Perfusion4
Adults, Pediatrics
±2 digits
Neonate
±3 digits
Carboxyhemoglobin Saturation Accuracy (%SpCO)
5
1% - 40% ±3 digits
Methemoglobin Saturation Accuracy (%SpMet)
5
1% - 15% ±1 digit
Pulse Rate Accuracy
Pulse rate:
25 - 240 bpm
No Motion1
Adults, Pediatrics, Neonate
±3 digits
Motion3
Adults, Pediatrics, Neonate
±5 digits
Low Perfusion4
Adults, Pediatrics, Neonate
±3 digits
Resolution
Saturation (%SpO2)
1%
Carboxyhemoglobin saturation (%SpCO), digital display
1%
Methemoglobin saturation (%SpMet), digital display .
1%
Pulse Rate (bpm)
1 bpm
Trending
72 hours of trending at 2 second resolution
Mode
Averaging mode:
2, 4, 8,10, 12, 14 or 16 seconds8
Sensitivity:
Normal and Maximum
9
and APOD
Alarms
Audible and visual alarms for high/low saturation and pulse rate (SpO2 range 1-100%, pulse rate
range 25-240 bpm, SpCO range 1%-99%, SpMet range 1%-99.9%)
Sensor condition, system failure and low battery alarms
Alarm Volume: 70 dB (min)
Display/Indicators
Data display: %SpO2, %SpCO, %SpMet, pulse rate, pleth waveform, alarm status, trends, status
messages, Signal IQ, perfusion index, pleth variability index, APOD and FastSat
Display update rate:
1 second
Response Time:
< 20 second delay
© 2013 DanMedical Ltd
95